Page 8 - BioChemFluidics 2012 Catalog 70p

Basic HTML Version

7
075P Series PINCH VALVES continued
C. 3-way valves (one port normally open, other port
normally closed)
2
PART NO.
INNER
DIAMETER
inch/mm
OUTER
DIAMETER
inch/mm
MIN
PRESSURE
inHg vac.
MAX
PRESSURE
psi/bar
POWER
AT 21°c
Watts
12 VDC valves with silicone tubing, two 6” (152mm) lengths of
silicone tubing joined by ‘Y’ connector installed in the valve.
075P3MP12-23S
0.023 /
0.6
0.093 /
2.4
28.2
20 /
1.3
3.6
075P3MP12-01S 1/32 / 0.8 3/32 / 2.4 28.2
15 /
1.0
3.6
075P3MP12-50S
3
0.050 /
1.3
0.134 /
3.4
28.2
15 /
1.0
3.6
075P3MP12-02S 1/16 / 1.6 1/8 / 3.2 28.2
15 /
1.0
3.6
12 VDC valves with C-Flex® tubing, two 6” (152mm) lengths of
C-Flex® tubing joined by ‘Y’ connector installed in the valve.
075P3MP12-23B
0.023 /
0.6
0.093 /
2.4
28.2
20 /
1.3
3.6
075P3MP12-01B 1/32 / 0.8 3/32 / 2.4 28.2
15 /
1.0
2.9
075P3MP12-02B 1/16 / 1.6 1/8 / 3.2 28.2
15 /
1.0
2.9
D. Available options, 075P series pinch valves
1.
“Quiet” operation (suffix = “Q”)
. The quiet option reduces the
valve noise level from about 72 dB (A) to less than 50 dB (A). The
quiet option is available for:
a. 075P2NC 2-way normally closed valves
b. 075P2NO 2-way normally open valves
The quiet option cannot be used on 075P3MP 3-way valves. The
noise level of the 3-way valves is inherently attenuated by the
tubing which dampens the motion in both directions.
2.
Mounting holes (suffix = “M”)
. Threaded holes situated on the
bottom of the valve body. Refer to installation drawing (page 8) for
details and dimensions.
The mounting holes cannot be used in combination with the
manual override button.
3.
Position feedback sensor (suffix = “F”)
. The feedback sensor
provides secondary, independent confirmation of the plunger
position (more details on page 14). The feedback option is available
for:
a. 075P2NC 2-way normally closed valves
b. 075P3 3-way valves
4.
Factory-installed mounting flange (suffix = “L”)
. The mounting
flanges are available for all 075P pinch valve configurations. They
will be installed at a 90º angle to the tubing with the orientation to
the outside of the bulkhead.
This mounting flange can be purchased as a stand-alone item under
part number MU-075 (more details on page 16).
E. “Hit and Hold” voltage reduction
Within 100 ms of valve actuation, the voltage required to hold the
valve in position can be reduced. The voltage step-down results in
the dual benefits of power conservation and heat reduction. The
amount of the voltage reduction possible is 66% (i.e. 24 Vdc to 8 Vdc
or 12 Vdc to 4 Vdc) for all 075P pinch valves that do not incorporate
the “quiet” feature. This results in a power usage reduction of 89%,
resulting in a correspondingly large drop in heat generation.
For the 075P pinch valves with the “quiet” feature, voltage can be
reduced by 50% (i.e. 24 Vdc to 12 Vdc and 12 Vdc to 6 Vdc) after 100
ms for a power usage drop of 75%.
This voltage reduction can be accomplished using the Bio-Chem
Fluidics CoolCube™ hit-and-hold circuit (more details on page 14).
NOTES:
1. For tubeless version use “#” in place of “S”
2. The operation of 3-way valves can be compromised by extended continuous
operation. A “Hit and Hold” circuit is strongly recommended to reduce heat generation
in extended use applications.
3. 075P3MP12-50S Pinch Valve is not molded
Option Highlight
Quiet Option
PTFE disc
Armature
Body
Features: Reduces the valve noise level from about
72 dB (A) to less than 50 dB (A).
How it works: A PTFE disc is inserted between the
internal components. The disc cushions the motion
of the plunger, leading to a substantial reduction in
noise.
Availability: 075P2NC, 075P2NO, 100P2NC and
100P2NO valves.
Ordering info: Add suffix “Q” to part number.