UNIQUE FEATURES
150 VOCs Analysi s
HARDWARE:
Switch power supply
Advanced temperature control
Compact PCB board
SOFTWARE :
QAlQC
print out end-user friendly
QAlQC
reports listed in a rolling calendar
o
Dynamic gas flow display and flow direction for all sample paths
NUTECH 8900 SPECIFICATIONS
o
Three stage cryogenic traps
Glass bead : -190 "C to 250 "C
o
Tenax mult imedia trap: -19WC to 250 "C
o
Cryofocuser : -190 "C to 250 "C
Sensitivity and detection range
o
0,1ppbv to percent level for most VOCs
Concentration ratio:
>
1000:1
o
Accuracy and repeatabi lity
o
Within 10% RSD for most VOC compounds with a sample
Loading range of 4- 1OOOml
o
Sample handling
All tubes are silica coated stainless steel
4-2000ml loading range with 0-100% relative humidity
o
MFC operating range 5- l20mllmin with +/-2% accuracy with an opt ional
customized flow range
Temperature contro l: +/-2 'C accuracy
Gas inlets: independen t internal standard inlet and 3 additional samp le inlets 16 sample inlets with optional autosampler
Operating environment: I IOv/60hz or 220v/50hz with max l500w, 0-40"C/90% RH operating range ,
Dimensions: l4"(W) X20"(H) X20"(D)
Weight: 65 lbs.
Headqu a rt er s: GD Env ir onme nta l Sup plies , Inc.
Sales Office: New Star Environmental, Inc,
651 N. Pl ano Rd. #429, Ri chardson TX 75081
3293 Ashbur ton Chase NE, Roswell GA 30075
6
E-mai l: drdai@gd air.com
E-ma il: Sa les@NewSla rE nviro nmenl al.com
Web sit e: www. nu tec hins .com, T: 972 480 8908; F: 972 480 8308
Websile: www.NewSlarEnvironm enlal.com. T: 770998 0296: F: 7709989794