ColumnPerformance
after 400Cycles
After 400 cycles to 430
0
o
o
C, aPolywax1000 sample gave
a column resolutionvalue of 2.7
Amixed sample of theD2887 standard andPolywax
1000was injected to calibrate the column forC10
throughC100.
Adiluted sample of Pennsylvania light crude oilwas then
analyzed and compared to the calibrationmixture.
Adequate resolutionof the hydrocarbons in the crude oil
samplewas obtained even though the columnwas below
theminimum resolution criteria of themethod.