MRC-2013-Catalog - page 648

648
ULTRASONIC PROCESSORS
U
Digital
Ultrasonic Probe
Stand
Small Bottle
Optional CUP:
Parameters:
Power Rating:
150 -2000 w
Operating Frequency:
20-22 KHZ
Input power:
100 -240V, 50 - 60Hz
Programmable Timer:
0-999 minutes
Crusher capacity:
0.5-1200 ml (expansion available)
99 program can be stored.
Optional:
Standard Probes -
Probes are made from titanium
and machined to specific sizes and shapes. When
driven at their resonant frequency, they expand and
contract longitudinally. This mechanical vibration is
amplified and transmitted down the length of the
probe. In liquid, the probe causes cavitation which
constitutes the main mechanism for sample
processing. Choosing the appropriate horn is
extremely important. The sample volume to be
processed is directly related to the tip diameter.
Cup horn/probe -
Cup horn for ultrasonicator :
This indirect Sonication eliminates the need for a
probe to come in contact with your sample.
This technique is often described as a high intensity
ultrasonic bath. The ultrasonic energy is transmitted
from the horn, up through the water and into a
vessel or multiple sample tubes.
Indirect sonication is most effective for very small
samples because foaming and sample loss are
eliminated. Pathogenic or sterile samples are ideal
for this method because aerosols and cross
contamination are prevented.
The Cup Horn and Microplate Horn deliver indirect
sonication and are ideal for many high throughput
applications.
Microtube holder -
up to 2-8 samples (0.1-2ml) can
be treated simultaneously. Pressure plate holds tubes
in place. Avoid contamination by toxic samples.
continuous cell flow -
316 stainless steel with double
walls through which cooling water can circulate,
This flow cell allows for 10 L per hour sonication.
It works with 10-20mm probes; and with 950w to
2000w syclon ultrasonic processors.
Optional Glass Cell Flow:
Optional Probes:
Ordering Information:
Ultrasonicator includes:
Generator/transducer
Convertor with cable
One replaceable probe
Power cable
Wrench set
Operation manual
Please specify desired voltage for export.
Sonicating
liquid recycle
Chilling water
recycle to cool
the sonicating
liquid
Probe
Liquid
flow out
Chilling
water in
Liquid
flow in
Chilling
water out