MRC-2013-Catalog - page 259

259
F
Automatic / Semi Automatic
The instrument is designed and made as per GB/T261 “Test Methods for Flash Point of Petroleum Products (Closed Cup
Methods)”. It is used to determine the lowest temperature of a petroleum product sample when the mixture of sample
vapor and air around the sample contacts flame and flashes when heated in a closed cup under stipulated condition.
The lowest temperature is the flash point of the petroleum product sample. The instrument has function of automatic
program temperature controlling, automatic barometric pressure modification, and modified value calculation. It can
automatically control measurement procedures and test data, and it can also save, analyze and print out test data.
FP-26A, Automatic PMCC Flash Point Tester
FP-26A
The instrument is designed and made as per the National Standard of People’s Republic of China GB/T5208 “Test
Methods for Flash Point (Rapid Equilibrium Closed Cup Method)”. It can be used in the field of petroleum,
chemistry, paints, railway, aviation, power plant, inspection, military, and scientific research to determine flash point of
petroleum products rapidly.
The instrument is suitable to make closed cup flash point determination for paints, adhesives, solvents, petroleum, and
relative petroleum products, the flash points of which are in the range of 30°C
50°C. It can also meet the
requirements of ISO 1523 and ISO 3679.
Main technical characteristics
It is designed as per GB/T5208 (Test Methods for Flash Point (Rapid Equilibrium Closed Cup Method)”. It can
determine closed cup flash point at a low temperature rapidly.
It adopts semiconductor refrigeration device and cooling water refrigeration device. It is small, convenient and with
rapid refrigeration speed.
It needs fewer samples. The sample for each test is only 2 ml (For solid or semisolid sample, it is 4 ml);
The test can be automatically completed excepting flame ignition. The test data can be automatically printed.
FPT-5208, Rapid Low Temperature Closed
Cup Flash Point Tester
Model
FP-26A
Power supply
AC 220 V±10%, 50 Hz
Temperature measurement
Measurement range
ambient temperature -250°C
Flash point
lower than 104°C, error is 2°C
higher than 104°C, error is 6°C
Resolution
0.1°C
Precision
0.5%
Basic parameters
Temperature rising rate
as per GB/T261
Ignition mode
gas flame is 3-4llim
Ambient temperature
10~40°C
Relative humidity
80%
Total power consumption
not more than 350W
Model
FPT-5208
Power supply
AC 220 V±10%, 50 Hz
Measurement range
-30°C~50°C
Measurement accuracy
Repeatability
less than 2°C
Reproducibility
less than 2°C
Temperature controlling
accuracy
±0.5°C
Igniter
an electron igniter
Refrigeration type
It uses semiconductor for re-
frigeration (It connects to outer
cooling water)
Total power consumption
not more than 300 W
Ambient temperature
5~30°C
Relative humidity
30~80%
Total power consumption
not more than 350W
FPT-5208
FLASH POINT