MRC-2013-Catalog - page 182

182
D
DRY BATHS
Max. 100°C/150°C, 3 Blocks
Model
DBD-003
DBD-003E
DBA-001
Overall dimension(mm)
370x245x135 290x290x110 210x235x115
Net weight
(Not include blocks)
3.9kg
6.0kg
2.1kg
Temp. setting
Digital (3-digit)
Digital (3-digit)
Analog
Temp. reading
Digital LED
Digital LED thermometer
Temp. range
Ambient to
99.9°C
Ambient to
150°C
25-105°C
Temp. accuracy
±0.1°C
±0.1°C
±0.2°C
Uniformity (at 37°C)
±0.2°C
±0.2°C
±0.5°C
Temp. sensor
PT-100
K-Type
*NTC
Temp. control
PID
PID
PID
Overheat alarm
Yes
Yes
No
Heating element
450W
450W
200W
Timer
Digital 99hr59min Digital 99hr59min
No
Insulation
No
Yes
No
DB-Series, Dry Bath Incubator
DBA-001
DBD-003
DBD-003E
*NTC: Negative Temperature Coefficient
Features:
Applications for incubating DNA specimens, cross matching and screens in
blood bank, incubating melted agar, Enzyme assays, digestions, Residue
tests for milk & also great for serology, hematology, pathology and
biochemistry etc.
Choose from various heating blocks that accommodate test tubes,
cuvettes, and micro centrifuge tubes.
Each block has a well to accommodate a removable thermometer,
which monitors temperature of blocks.
When solid block (no holes) is used, it can function as HOT PLATE.
Application:
Solid anodized aluminium blocks resist
the hardest environments and provide
maximum heat retention.
All blocks include thermometer well for
easily measuring block
temperature.
Number
Tube size
Openings No. /
Ø
302-1-111101
6mm
30/6.5mm
302-1-111112
10mm
20/10.5mm
302-1-111113
12 or 13mm
20/13.5mm
302-1-111114
15 or 16mm
12/16.6mm
302-1-111115
18mm
12/18.6mm
302-1-111116
20mm
12/20.7mm
302-1-111117
25mm
6/25.7mm
302-1-111118
1.5ml Eppendorf
20
302-1-111119
0.5ml Eppendorf
20
302-1-111110
25+12+6mm or your request
3+5+6 holes
302-1-111111
Solid (no holes) used as HOT PLATE
302-1-111112
Microplate 96 wells
Block size
W100xD75xL50(mm), W155xD100xL50(mm) Microplate
DBD-003 & DBD-003E
Activate auto-off function to automatically shut out heater at the end of
timed cycle.
Digital temperature display for actual temperature PV and setting Sv.
Timer start working when the preset temperature has been stabilized
ensures an accurate incubation effect.
When complete the heat cycling, the timer will alarm and display “END”.
Memory stores the set temperature and time of previous operation, use
these set points to repeat the new operation and save time, or preset your
new desired set points.
DB-006E-150 features double wall construction with insulator by fiberglass for durability and good heat retention.
Maximum temperature is designed up to 150°C capable of functioning more applications.
Safety devices: Over temperature alarm; Heating failure alarm; Double overheating protection.
1...,172,173,174,175,176,177,178,179,180,181 183,184,185,186,187,188,189,190,191,192,...738
Powered by FlippingBook