182
D
DRY BATHS
Max. 100°C/150°C, 3 Blocks
Model
DBD-003
DBD-003E
DBA-001
Overall dimension(mm)
370x245x135 290x290x110 210x235x115
Net weight
(Not include blocks)
3.9kg
6.0kg
2.1kg
Temp. setting
Digital (3-digit)
Digital (3-digit)
Analog
Temp. reading
Digital LED
Digital LED thermometer
Temp. range
Ambient to
99.9°C
Ambient to
150°C
25-105°C
Temp. accuracy
±0.1°C
±0.1°C
±0.2°C
Uniformity (at 37°C)
±0.2°C
±0.2°C
±0.5°C
Temp. sensor
PT-100
K-Type
*NTC
Temp. control
PID
PID
PID
Overheat alarm
Yes
Yes
No
Heating element
450W
450W
200W
Timer
Digital 99hr59min Digital 99hr59min
No
Insulation
No
Yes
No
DB-Series, Dry Bath Incubator
DBA-001
DBD-003
DBD-003E
*NTC: Negative Temperature Coefficient
Features:
●
Applications for incubating DNA specimens, cross matching and screens in
blood bank, incubating melted agar, Enzyme assays, digestions, Residue
tests for milk & also great for serology, hematology, pathology and
biochemistry etc.
●
Choose from various heating blocks that accommodate test tubes,
cuvettes, and micro centrifuge tubes.
●
Each block has a well to accommodate a removable thermometer,
which monitors temperature of blocks.
●
When solid block (no holes) is used, it can function as HOT PLATE.
Application:
●
Solid anodized aluminium blocks resist
the hardest environments and provide
maximum heat retention.
●
All blocks include thermometer well for
easily measuring block
temperature.
Number
Tube size
Openings No. /
Ø
302-1-111101
6mm
30/6.5mm
302-1-111112
10mm
20/10.5mm
302-1-111113
12 or 13mm
20/13.5mm
302-1-111114
15 or 16mm
12/16.6mm
302-1-111115
18mm
12/18.6mm
302-1-111116
20mm
12/20.7mm
302-1-111117
25mm
6/25.7mm
302-1-111118
1.5ml Eppendorf
20
302-1-111119
0.5ml Eppendorf
20
302-1-111110
25+12+6mm or your request
3+5+6 holes
302-1-111111
Solid (no holes) used as HOT PLATE
302-1-111112
Microplate 96 wells
Block size
W100xD75xL50(mm), W155xD100xL50(mm) Microplate
DBD-003 & DBD-003E
●
Activate auto-off function to automatically shut out heater at the end of
timed cycle.
●
Digital temperature display for actual temperature PV and setting Sv.
●
Timer start working when the preset temperature has been stabilized
ensures an accurate incubation effect.
●
When complete the heat cycling, the timer will alarm and display “END”.
●
Memory stores the set temperature and time of previous operation, use
these set points to repeat the new operation and save time, or preset your
new desired set points.
●
DB-006E-150 features double wall construction with insulator by fiberglass for durability and good heat retention.
Maximum temperature is designed up to 150°C capable of functioning more applications.
●
Safety devices: Over temperature alarm; Heating failure alarm; Double overheating protection.