Restek-OEM-Parts_780p - page 59

Super-Clean™GasFilters
• High-purityoutput ensures 99.9999%pure gas.
• “Quick connect”fittings for easy, leak-tight cartridge changes.
• Glass inside toprevent diffusion; plastic outside for safety.
Designed for fast, simplecartridgechanging
Cartridge systemsmake changing gas filters quick and easy, andSuper-Clean™ gas filters are the latest in
cartridge-style gas filtration.A baseplate allows cartridges to be exchangedwithout introducing oxygen.
Spring-loaded checkvalves sealwhena filter is removedandopenonlywhenanew filterhasbeen locked
inplace.There is no longer aneed for loosening and tightening fittings every time a trap is changed, and
your systemwill not become contaminatedduring the process.
With available 2- or 3-position baseplates, you can purify all GC gas streams at one location. Figure 1
shows some possible filter cartridge combinations using these baseplates. Any combination is possible
because anyRestek Super-Clean™ filter cartridge canbeusedwith anyRestekbaseplate.
High-purityoutput for improvedsensitivity (Table I)
TheTripleFilter cartridge (cat.#22020) is ideal forpurifyingcarriergas.This trapcontainsoxygen,mois-
ture,andhydrocarbon scrubbers inonecartridge.Thepurityof your carriergasafter flowing through the
TripleFilter is better than six-9s (99.9999%pure),which is ideal for sensitivemass spectrometry (MS) or
ECDanalyses, and forprotectingyour analytical columns against damage from contaminated carrier gas.
The Fuel Gas Filter cartridge (cat.# 22022) is perfect for purifying flame ionization detector (FID) fuel
gases, removingbothmoistureandhydrocarbons.Using theFuelGasFilter forFIDhydrogenandairwill
produce a stable baseline, improvingoverall reproducibility and sensitivity.
Moisture
cat.# 22028
Oxygen
cat.# 22029
Hydrocarbons
cat.# 22030
Fuel Gas
1
cat.# 22022
Triple
2
cat.# 22020
Helium Specific
2
cat.# 21982
>99.9999
>99.9999
>99.9999
>99.9999
>99.9999
>99.9999
11 bar
159psi/
7 L/min.
11 bar
159psi/
7 L/min.
11 bar
159psi/
7L/min.
11 bar
159psi/
7L/min.
11 bar
159psi/
7L/min.
11 bar
159psi/
7L/min.
Inert carrier gas
Air
Hydrogen
Inert carrier gas
Inert carrier gas
Air
Hydrogen
Inert carrier gas
Air
Hydrogen
Inert carrier gas
Helium
Yellow to clear
Green to grey
No indicator
Yellow to clear
Yellow to clear
Green to grey
Yellow to clear
Green to grey
7.2
NA
NA
3.5
1.8
1.8
1000
500
500
12
3
24
3
4
3
>2
>2
>2
>1.5
>1
>1
Type of Filter
Outlet
Gas Quality (%)
Maximum
Pressure/
Maximum
FlowRates
Use for:
Indicator
Color Change
Capacity
H
2
0 (g) O
2
(mL) Hydrocarbons (g)
Estimated
Lifetime
(years)
carrier gas
hydrogen
Triple Filter
hydrocarbons/
moisture/oxygen
Fuel Gas Filter
hydrocarbons/
moisture
air
Fuel Gas Filter
hydrocarbons/
moisture
carrier gas
Triple Filter
hydrocarbons/
moisture/oxygen
hydrogen
Fuel Gas Filter
hydrocarbons/
moisture
air
Fuel Gas Filter
hydrocarbons/
moisture
CompleteFilter BundleKit for FID
BurnerGasKit for FID
TripleKit for
MS, ECD, NPD
Figure1
Filter cartridges canbeconfigured fordifferent applications.
Table I
EachSuper-Clean™ filterprovideshigh-purityoutletgas.
cat.# 22020
cat.# 22022
cat.# 22022
cat.# 22020
cat.# 22022
cat.# 22022
1
Removes hydrocarbons,
moisture.
2
Removes hydrocarbons,
moisture, oxygen.
3
As
n
-butane.
didyou
know
?
All Super-Clean
filter
cartridges (except
hydrocarbon filter cat.#
22030) feature easy-to-read
indicators. The indicator
code is shown on every trap
so there is no confusion
about when to replace it.
SUB-SECTION TITLE
Super-Clean™ Gas Filter System
800-356-1688
20
tech
tip
Oxygen andMoisture Traps
We highly recommend
oxygen andmoisture traps for
make-up gaswhen operating
sensitive detectors such as
electron capture detectors
(ECD).The hydrogen reaction
gas for sensitive electrolytic
conductivity detectors (ELCD)
also requires a hydrocarbon
trap, to remove trace
impurities.