Fittings-Chromatography-2014 - page 403

Flowmaster
®
Ideal forHeavy-DutyProcesses
XX
Ideal for dispensing and filling applications in adusty, humid
or corrosiveenvironment and in clean room areas
XX
Protection ratingof IP 65
Optimized for IncreasedHygienicRequirements
XX
Stainless steel housing
XX
Tube-loadingunder sterile conditionswithout aspirating air
XX
Easydisassemblyof thepumphead
XX
Thorough cleaning thanks toeasydisassembly and
reassemblyof thepumphead
Safety
XX
PumpStopsWhenOpening theTube-Bed
XX
MultipleOverloadProtection
FlowmasterFMT300
37mL/min–13L/min
XX
1 channel
XX
3 convex stainless steel rollers
XX
Automatic tube retention
XX
Standard tubing 6.4–15.9mm ID, wall thickness 3.2mm, differential
pressuremax. 2bar (30psi)—depends on tubingmaterial used
PartNo.
Description
Flow rates
mL/min
Channels Rollers Speed rpm
FLOWMASTER
ISM1020A
Flowmaster FMT300
230V50Hz
37–13,000
1
3
5–500
ISM1022A
Flowmaster FMT300
115V60Hz
37–13,000
1
3
5–500
ACCESSORIES
IS10279
Foot switch, seepage109
RELATEDPRODUCTS
Accessories
Page 109
Tubing
Pages 75–89
XX
Insert theTube (Easily andFast)
XX
PressDown the Lever
(AutomaticallyCorrect Pressure
Settingof theTube)
XX
Start thePump!
TUBEEXCHANGE IN 5 SECONDS!!
TIP
FLOWRATES&TUBING
Tubing Information
FlowRates in L/min
Tygon
®
E-LFL
PartNo.
PharMed
®
PartNo.
Wall
(mm)
Tubing
ID (mm)
rpm
5
rpm
10
rpm
50
rpm
100
rpm
200
rpm
300
rpm
400
rpm
500
SCE0393 MF0015
3.2 6.4 0.037 0.074 0.37 0.74 1.5 2.2 3.1 3.7
SCE0395 MF0016
3.2 9.5 0.08 0.16 0.80 1.6 3.2 4.8 6.4 8.0
SCE0396 MF0034
3.2 12.7 0.10 0.20 1.0 2.0 4.0 6.0 8.0 10.0
SC0696
3.2 15.9 0.13 0.26 1.3 2.6 5.2 7.8 10.4 13.0
Approx. values: determinedwithwater, at 22 °C, nodifferential pressure, PharMed tubing.
SPECIFICATIONS&DETAILS
Motor type
DCmotor
Speed
5 to500 rpm
Speed setting
Resolution 0.1 rpmmembrane key-pad, LEDdisplay
Power consumption
500W
Mains connection
230VAC/50Hz,115VAC/60Hz, selectable
Protection rating
IP 65
Depth/Width/Height
500 x 220 x 262mm
Weight
26 kg
Interfaces
PLCcompatible interfacewith status information forprocess control
systems (the level of the inputs canbeconfigured: 5, 12, or 24V).
XX
Speed control
(0–5or 0–10V,
0–20or 4–20mA)
XX
Start/stop,
rotationdirection
XX
Autostart
XX
Speedoutput
XX
Digital output (potential free)
(error, okay, busy)
Settingsmenu
XX
Configurationof analog
interface
XX
Entryof basic settings, e.g.
rpm, time, etc.
XX
Foot switch control
XX
Rotation speed (%or rpm)
XX
Service lifeof tubing
XX
Timer function, etc.
98
LABORATORYPUMPS
Peristaltic Pumps
1...,393,394,395,396,397,398,399,400,401,402 404,405,406,407,408,409,410,411,412,413,...581