Restek ESSENTIALS 2010-2012 - page 171

Superior resolutionandpeak shapeonMXT®-1HTSimDist columns
result inmoreaccurate final boilingpointdeterminations.
Column:
MXT
®
-1HT SimDist, 5m, 0.53mm ID,
0.20μm (cat.# 70115)
Sample:
C5-C100, 1% in carbon disulfide
Inj.:
1μL on-column (PTV)
Inj. temp.:
53°C to 430°C@ 10°C/min. (hold 5min.)
C50 skew= 0.77
C50/C52 R= 2.5
C100
Bleed= 9pa
LowerBleed
SimDist=Longer
Lifetime!
Carrier gas: helium, constant flow
Flow rate:
18mL/min.
Oven temp.: 50°C to 430°C@
10°C/min. (hold 5min.)
Det.:
FID@ 430°C
Instrument: Shimadzu 2010
0
10
20
30
40min.
• Stableup to430°C—lowestbleedcolumn, for longer lifetime.
• MeetsallASTMD6352andD7500specifications.
• 100%dimethylpolysiloxanephaseallowseasycomparisons
tohistoricaldata.
• Unbreakablecolumnsoffer securitywhenusinghydrogen.
MXT®-1HTSimDistColumn
(Siltek® treated stainless steel)
ID df (μm) temp. limits
length cat. #
0.53mm 0.10 -60 to400/430°C 5-Meter 70112
0.53mm 0.20 -60 to400/430°C 5-Meter 70115