innovativepharmaceutical solutions
Pharmaceutical
Rxi®-624SilMS
The
GoTo
GCColumn forFast, EffectiveVolatile
ImpuritiesMethodDevelopment
GoTo…
the right column first
Forbetter retentionof polar analytes and improvedaccuracy,peak shape, and response for activecompounds.
GoTo…
mass specdirectly
With the lowest bleed 624 column available; stable up to 320 °C, for easy transfer of methods toGC/MS.
GoTo…
thenext batch faster
With the best-in-classG43 forUSPmethods.
ZB-624
Integrationnot possible
below25ng
Rxi®-624SilMS
5-100ngon-column,
R
2
=0.99996
Exceptionalpeaksymmetry
and retention timestability
result inhighsensitivityand
linearityon theRxi®-624SilMS
Column: 30m, 0.32mm ID, 1.8µm (cat.# 13870); Sample: Isopropylamine; Diluent: DMSO; Conc.: 100µg/mL; Injection: Inj. Vol.: 1µL split (split ratio
20:1); Liner: 5mm SingleGooseneckwithWool (cat.# 22973-200.1); Inj.Temp.: 250 °C; Oven Temp: 50 °C (hold 1min.) to 200 °C at 20 °C/min. (hold 5min.);
Carrier Gas: He, constant flow; Linear Velocity: 37 cm/sec.; Detector: FID@ 250 °C Instrument: Agilent/HP6890GC
Theversatilityof an
Rxi®-624SilMS
columnmakes ita
perfect fit for
QualityByDesign.
Rxi®-624SilMSColumns (fusedsilica)
• Lowbleed, high thermal stabilitycolumn—maximum temperatures
up to320°C.
• Inert—excellentpeak shape for awide rangeof compounds,
includingacidicandbasiccompounds.
• Selective—highlyselective for residual solvents, greatchoice forUSP<467>.
• Manufactured for column-to-column reproducibility—well-suited for
validatedmethods.
ID df (µm)
temp. limits
20-Meter
30-Meter
60-Meter
0.18mm 1.00 -20 to 300/320°C 13865
0.25mm 1.40 -20 to 300/320°C
13868
0.32mm 1.80 -20 to 300/320°C
13870
13872
0.53mm 3.00 -20 to 280/300°C
13871
TECHNOLOGY
3-IN-1
(midpolarityCrossbond® silarylenephase; equivalent to6% cyanopropylphenyl/94%dimethyl polysiloxane)