Chromatography Columns 2013+ - page 186

129
GCCOLUMNS |
PACKED/MICROPACKEDCOLUMNS
Specialty Packed Columns
Figure3
Refinerygas calibration standardonaRestek
refinerygaspackedcolumn set.
Figure2
OPNonRes-Sil®Cpackinghasunique
selectivity for
cis
-2-buteneand1,3-butadiene.
Retention
Indices
1. methane
100
2. ethane/ethylene
200
3. acetylene
260
4. propane
300
5. propylene
321
6. propadiene
345
7. isobutane
386
8. butane
400
9. 1-butene
422
10. isobutylene/
trans
-2-butene
434
11.
cis
-2-butene
443
12. 1,3-butadiene
454
13. isopentane
488
14. pentane/3-methyl-1-butene
503
15. 1-pentene
522
16.
trans
-2-pentene
533
17.
cis
-2-pentene
540
18. 2-methyl-2-butene
549
1
3
5 2
6
7 8
9
10
11
12
13
14
15 16
17 18
min. 0
5
10
4
OPN on Res-Sil
®
C, 80/100mesh, 12' x 2mm ID x
1
/
8
" OD in SilcoSmooth
®
tubing (cat.# 80437). 20µL on-column injection of refinery gas.
Concentration:
0.1-6 absolutemole%
Oven temp.:
50°C
Inj. & det. temp.: 200°C
Carrier gas:
helium
Flow rate:
30mL/min
Reference standard courtesy of ACAnalytical Controls, Bensalem, PA.
GC_PC00373
cis
-2-butene/1,3-butadiene
resolved tobaseline
1
min.
10
20
30
1. C5=C6+
2. methane
3. ethane
4. ethylene
5. propane
6. acetylene
7. propylene
8. cyclopropane
9. isobutane
10. propadiene
11.
n
-butane
12. 1-butene
13. isobutene
14.
trans
-2-butene
15.
cis
-2-butene
16. 1,3-butadiene
17. isopentane
18.
n
-pentane
3,4 5
2
6
7
8
9
10
11
12
13
14
15
16
17 18
2abc Refinery Gas Column Set (cat.# 88000-875) (3 column set)
Oven temp.:
60°C
Inj. temp.:
150°C
Det. temp:
150°C FID
Flow rate:
30mL/min., helium
Sample:
refinery gas
Sample size:
1cc
GC_PC00374
Figure1
n
-OctaneonRes-Sil®Cpackinghasunique
selectivity forunsaturated lighthydrocarbons.
for
more
info
See
page135
formore informationonRes-Sil®packingmaterials.
1
min. 0
2
4
6
1. methane
2. ethane
3. ethylene
4. acetylene
5. propane
6. propylene
7. isobutane
8.
n
-butane
9. 1-butene
10. pentane
11. isobutylene
12. 1,3-butadiene
13. 1-pentene
14.
trans
-2-pentene
15.
cis
-2-pentene
16. 2-methylbutene-2
n
-octane 80/100 Res-Sil
®
C
20',
1
/
8
" OD x 2mm ID, SilcoSmooth
®
tubing (cat.# 80436)
Oven temp.:
60°C
Inj. temp.:
150°C
Det. temp:
150°C FID
Flow rate:
30mL/min. He
Sample:
refinery gas C1-C5
Sample size:
20µL
2,3
4
5
6
7 8
9
11
12
8
10
12
13
14
15
16
GC_PC00372
10
*Please add column instrument configuration suffix number to cat.#when
ordering. See page 143.
**This column set is for a valving system; therefore, packingmaterial is filled to
ends of columns.
SpecialColumns forUnsaturatedLightHydrocarbons
• Faster separations of C1 toC4hydrocarbons.
• Res-Sil®packing replaces Porasilmaterials.
n
-OctaneonRes-Sil®CPackedColumn
This packed columnhas unique selectivity for resolvingunsatu-
rated light hydrocarbons (Figure 1).
OPNonRes-Sil®CPackedColumn
This column separates the light hydrocarbons, and baseline
resolves
cis
-2-butene from1,3-butadiene (Figure 2).
2abcRefineryGasColumnSet
This 3-column set is finely tuned to resolve light hydrocarbons.
Whenused in theproper valving system, itwill eluteC5+hydro-
carbons aheadof C1 throughC4hydrocarbons. (Figure 3)
LightHydrocarbonAnalysis
Description
cat.#*
price
n
-Octane on Res-Sil C, 80/100 (20', 2.0mmID,
1
/
8
" SilcosmoothOD)
80436-
OPN on Res-Sil C, 80/100 (12', 2.0mmID,
1
/
8
" SilcosmoothOD)
80437-
2abc Refinery Gas Column Set (3 column set)**
88000-
Website :
E-mail :
TelNo : 03 9762 2034 . . . inAUSTRALIA
Mar 2011