001_TOC.qxp - page 59

600
GCCHROMATOGRAMS |
ENVIRONMENTAL
Diesel Range Organics (DRO)
Column:
Rtx
®
-5, 15m, 0.25mm ID, 0.25µm (cat.# 10220)
Sample:
500µg/mL Diesel Fuel Composite standard (cat.# 31093) inmethylene chloride
Inj.:
1.0µL splitless (hold 0.5min.), 4mm single gooseneck inlet liner (cat.# 20904)
Inj. temp.:
260°C
Carrier gas:
helium, constant pressure
Linear velocity:
76cm/sec.@ 50°C
Oven temp.:
50°C (1min.) to 330°C@ 20°C/min.
Det.:
FID@ 330°C
DieselFuel
Rtx®-5
GC_EV00669
conventional
20°C/min.
temperature program
DROMix
Rtx®-5
1. C10
2. C12
3. C14
4. C16
5. C18
6. C20
7. C22
8. C24
9. C26
10. C28
1
2
3
4
10 9 8
7
6
5
Column:
Rtx
®
-5, 30m, 0.53mm ID, 1.0µm (cat.# 10255)
Sample:
Direct injection of DROMix (cat.# 31064)
Conc.:
50ng/µL
Oven temp.:
40°C (hold 2min.) to 310°C
@ 12°C/min. (hold 10min.)
Inj./det. temp.:
280°C/310°C
Carrier gas:
helium@ 65cm/sec.
FID sensitivity:
8 x 10
-11
AFS
min.
5
10
15
20
25
GC_EV00069
ForBiodiesel chromatograms,
seepages644–646.
Website :
E-mail :
TelNo : 03 9762 2034 . . . inAUSTRALIA
Mar 2011
1...,49,50,51,52,53,54,55,56,57,58 60,61,62,63,64,65,66,67,68,69,...166