VP014-v026 - page 89

026
89
TECHNICAL
INFORMATION
MLSL or MLRL
Fully Assembled Panel-Mounted Luer
FSLLR
FTLLB
CCLR or CCR
LNS
Panel
2X Ø .128
2X Ø .125
CCLR Only
Ø .255
.100
Ø .562
2X .128
2X .128
2X .188
.505
.505
Dimension A
Dimension A
Dimension A
Panel-mounted female luers, in both FTLLB (thread style) and FTLB (lug style) series, are shown beginning on page 40. When used
with our CCLR lock rings or CCR color coding rings and LNS lock nuts (see pages 41, 63 ), color coding of connections is easy.
The information provided here will help you mount the panel mount luers. Please note that they are intended for panel thickness of
1/16” to 1/8” (1.5 to 3 mm).
FTLLB series shown with MLSL luer, FSLLR lock ring, CCLR color coding lock ring, and LNS lock nut.
Threaded Panel Openings:
Tap the panel with threads from tube fittings, panel mounts or BPMBs and use a CCR color coded
ring. In this case, the fitting tightened into the panel prevents fitting rotation. Use an LNS lock nut for additional security; tighten
snug plus 1/4 turn only.
Non-Threaded Panel Openings:
Use any one of the three panel patterns shown below with a CCLR lock ring and LNS lock nut.
Together, the panel opening and CCLR prevent the panel mount fitting from rotating. Tighten LNS snug plus 1/4 turn only.
TECHNICAL:
BPF Series Properties and Performance
TECHNICAL:
Panel Mount Instructions
Maximum cycles
Maximum axial tensile resistance
Maximum working pressure
50,000 connect-disconnect cycles
20 lbf.
60 psi, 4.14 bar
0.0
0.2
0.4
0.6
0.8
1.0
1.2
1.4
0
10
20
30
40
50
60
70
Flow Rate Q (USGal/min)
0.0
0.8
1.5
2.3
3.0
3.8
4.5
5.3
0.0
1.0
2.0
3.0
4.0
5.0
Flow Rate Q (Liters/min)
Pressure Vs. Flow Rate, Water Pressure (bar)
Pressure (psi)
Maximum leak resistance
Minimum Fitting ID
Minimum Cv
< .007 psi / 0.05 kPa / sec.
.113” / 2.9 mm
.39