AirMonitoring - page 94

NUTECH 3551 PERFORMANCE - REPEATABILITY
Instruments: Nutech 3551DS Preconcentrator with Agilent 6890/5973 GC/MS
Standard Gas: EPA TO-14 , 50 ppb
Run 7 replic ate independent experiments ofTO-14 standard analysis using 100 ml sample volume of standard gas at 100 ml/min
flow rate. The results are shown in the following table:
Target Compounds
Dichlorodifluoromethane
Chloromethane
Dichlorotetrafluoroethane
Ethene, chloro-
Methane, bromo­
Chloroethane
Trichloromonofluoromethane
Ethene , 1, l-dichloro
Ethane,
I ,I
,2-trichloro
Methylene chloride
Ethane , 1, l -dichloro-
Ethene , 1,2-dichloro-, (Z)
Bromochloromethane(SS 1)
Chloroform
Ethane, 1,1, l -trichloro­
Ethane , 1,2-dichloro­
Benzene
Carbon tetrachloride
Benzene, 1,4-difluoro(SS2)
Trichloroethylene
Ethane , 1,2-chloro­
Cis-l,3-dichloropropene
Trans-l ,3-dichloropropene
Toluene
Ethane , 1,1,2-trichloro­
Tetrachloroethylene
1,2-dibromoethan e
Benzene, chioro­
Ethylbenzene
Xylene(p+m)
Styrene
Benzene, 1,4-dimethyl-
Ethane ,
I ,
1,2,2-tetrachloro
Benzene, l -bromo-4-fluoro-(SS3)
Benzene , 1,2,3-trimethyl­
Benzene, 1,2,4-trimethyl­
Benzene , 1,2-dichloro-
Benzene, 1,3-dichloro-
Benzene , 1,4-dichloro-
Benzene, 1,2,4-trichloro-
I , I
,2,3,4,4-Hexachloro -l ,3-butadiene
Headquarters: GD Environmental Supplies, Inc.
651 N. Plano Rd . #429, Richardson TX 75081
E-mail:
Website: www. nutechlns.com, T: 972 480 8908 ; F: 972 480 8308
AVG
RPD%
0.305
0.0081
2.66
0.125
0.0028
2.23
0.305
0.0081
2.66
0.074
0.002
2.70
0.060
0.0023
3.87
0.037
0.0012
3.10
0.367
0.0083
2.27
0.154
0.0053
3.42
0.215
0.0075
3.45
0.069
0.0027
3.82
0.177
0.0027
1.51
0.110
0.0022
1.98
0.419
0.0069
1.65
0.272
0.0046
1.69
0.362
0.0046
1.26
0.216
0.0029
1.35
0.366
0.0065
1.77
0.400
0.0043
1.07
2.424
0.0255
1.05
0.201
0.0025
1.23
0.033
0.0009
2.77
0.071
0.0021
2.89
0.021
0.0005
2.57
0.257
0.0038
1.46
0.092
0.0011
1.20
0.169
0.0028
1.61
0.109
0.0026
2.37
0.219
0.0012
0.56
0.380
0.0038
1.00
0.310
0.0028
0.91
0.134
0.001
0.76
0.315
0.0019
0.61
0.190
0.0041
2.17
0.851
0.0098
1.14
0.316
0.0039
1.22
0.345
0.0046
1.33
0.263
0.0039
1.47
0.263
0.0046
1.75
0.239
0.0036
1.49
0.182
0.0052
2.84
0.221
0.0048
2.18